Delete memory
When the PROG key is pressed, the symbol ´SClr´ starts flashing. Only if you
now press the Clear key will the memory content be deleted
delete memory by pressing the key: ,
PROG
CLEAR
`S C L r´SF
6.5.4 Sleep mode
For long-term monitoring with more prolonged cycles (i.e. longer than two minu-
tes) the measuring instrument can be operated in sleep mode. In energy-sa-
ving sleep mode the measuring instrument is switched off after each measuring
point scan (please note when using sensors with power supply !) and switched
on again automatically after the cycle expires ready for the next measuring
point scan. In this way with just one battery 10000 to 30000 measuring point
scans can be performed (depending on the number of measuring points and
the conversion rate); for a cycle lasting 15 minutes this represents a useful life
of between 100 and 300 days.
To perform data recording in sleep mode carry out the following steps:
1. Enter a cycle lasting at least two minutes (s.6.3.2):
2. In the DISPLAY MODE function ´DM´ activated sleep mode:
DISPLAY MODE function ´DM´
Select by pressing the FUNCTION key (s. 5.2)
The symbol with three lines means that sleep mode is switched off.
To activate sleep mode the (s. 5.4)
Press keys PROG, ▲, PROG, PROG or ESC:
3. To start the meas. operation press the: START/
STOP
The meas. instrument switches off with the display:
4. In the specified cycle the measuring instrument
switches on automatically, performs one mea-
suring point scan, and then switches off again.
5. Terminate sleep mode
Switch the measuring instrument OFF by means
of the slide switch (1) and then ON again;
(measuring operation is still started). OFF - ON
6. To terminate the measuring operation press
the START/
STOP key
- - - d DM
S L P d DM
S L E E P
40 ALMEMO
2390-5
Measuring operations
Komentarze do niniejszej Instrukcji